Multi-Echo fMRI
During a Repetition (defined by the TR), we can collect the signal once, at a certain known TE, to obtain a timeseries (Single-Echo, SE)...
...or multiple times, at different TEs, to obtain n timeseries (Multi-Echo, ME)
|----- - - -|----- - - -|----- - - -|
Assuming monoexponential decay, $ T_2^{\star} $ decay is:
\( \Large S(x,t,TE_n) = \) \( \Large S_0(x,t) \) \( \Large e^{-TE_n R_2^{\star}(x,t)}\) \( \Large + n \)
- Coil sensitivity (T1 changes & instability, such as: movement, scanner noise, ...)
- Reciprocal of $ T_2^{\star} $, field inhomogeneity (including tissue oxygenation state, i.e. BOLD signal!)
- (Kundu et al., 2017)
- Depends on $ TE_n $: If $ TE = 0 $, we can regress out $ S_0 $ (Bright & Murphy, 2013)
Take home message n. 1
We can fit our data to a model to separate different sources
\[ S_{TE_n} = S_0 e^{-TE_n R_2^{\star}} + n \]
\[ \ln(S_{TE_n}) = \ln(S_0 e^{-TE_n R_2^{\star}}) \]
\[ \ln(S_{TE_n}) = \ln(S_0) -R_2^{\star} TE_n \]
\[ \begin{cases} \ln(S_{TE_1}) = \ln(S_0) -R_2^{\star} TE_1 \\ \ln(S_{TE_2}) = \ln(S_0) -R_2^{\star} TE_2 \\\ln(S_{TE_3}) = \ln(S_0) -R_2^{\star} TE_3 \end{cases} \]
GLM: \( Y = a_0 - a_1 TE \)
$\overline{S_{0_{(fit)}}}$ vs $\overline{T_{2_{(fit)}}^{\star}}$
Improved coregistration with anatomical!
\[\small S_{OC}(t)=\sum{S_{TE_n}(t) w(T^{\star}_2)}_n \]
\[ w(T^{\star}_2)_n = \frac{TE_n \cdot e^{-TE_n/T^{\star}_{2(fit)}}}{\sum {TE_n \cdot e^{-TE_n/T^{\star}_{2(fit)}}}} \]
In this way, spatial CNR is maximised and the signal can be recovered in areas of drop-out (Posse, 1999)
One subject from OpenfMRI ds000258
TR: 2.47s, TEs: 12,28,44,60 ms, Res: 3.75x3.75x4.4 mm
Preprocessing:
- Discard first 10 tps
- Motion computation
- Realignment
- Smoothing (6mm)
- Nuisance regression (avg. WM, avg. CSF, 6 DoF Motion) & Detrend (Fifth Order)
- Optimal Combination of the signal
- Coregistration, Normalisation to MNI 152
Seed-based CAPs, 9 mm ROI around MNI coord: 3, -52, 21 mm (PCC)
Take home message n. 2
We can Optimally Combine data to get less noisy timeseries!
\(\small S_{TE_n} = S_0 e^{-TE_n R_2^{\star}} + n\)\(\small \qquad\qquad \overline{S_{TE_n}} = \overline{S_0} e^{-TE_n \overline{R_2^{\star}}} \)
\[\scriptsize S_0 = \overline{S_0}+\Delta S_0 \qquad\qquad \overline{S_0}>>\Delta S_0 \] \[\scriptsize R_2^{\star} = \overline{R_2^{\star}}+\Delta R_2^{\star} \qquad\qquad \overline{R_2^{\star}}>>\Delta R_2^{\star} \]
\[\small S_{TE_n} = (\overline{S_0} + \Delta S_0)e^{-TE_n(\overline{R_2^{\star}}+ \Delta R_2^{\star})} \]
\[\small S_{TE_n} = \overline{S_0} e^{-TE_n\overline{R_2^{\star}}}e^{-TE_n \Delta R_2^{\star}} + \Delta S_0 e^{-TE_n\overline{R_2^{\star}}}e^{-TE_n \Delta R_2^{\star}} \]
\[\small S_{TE_n} = (\overline{S_{TE_n}} / \overline{S_{TE_n}}) ( \overline{S_0} e^{-TE_n\overline{R_2^{\star}}} + \Delta S_0 e^{-TE_n\overline{R_2^{\star}}}) e^{-TE_n \Delta R_2^{\star}} \]
\[\small S_{TE_n} = \overline{S_{TE_n}}( 1 + \Delta S_0 / \overline{S_0}) e^{-TE_n \Delta R_2^{\star}} \]
\[\small S_{TE_n} = \overline{S_{TE_n}}( 1 + \Delta S_0 / \overline{S_0}) (1 -TE_n \Delta R_2^{\star}) \]
\[\small S_{TE_n} = \overline{S_{TE_n}} \big( 1 + \frac{\Delta S_0}{\overline{S_0}}\big)(1 -TE_n \Delta R_2^{\star}) \]
\[\small S_{TE_n} = \overline{S_{TE_n}} \big( 1 -TE_n \Delta R_2^{\star} + \frac{\Delta S_0}{\overline{S_0}} - \frac{\Delta S_0 TE_n \Delta R_2^{\star}}{\overline{S_0}} \big) \]
\[\small \frac{S_{TE_n}-\overline{S_{TE_n}}}{\overline{S_{TE_n}}} = -TE_n \Delta R_2^{\star} + \frac{\Delta S_0}{\overline{S_0}} \]
\[\small \Delta\rho=\Delta S_0 / \overline{S_0} \]
\[\small \Delta\kappa=\Delta R_2^{\star}\overline{TE} \]
\[\small S_{SPC} \approx \Delta\rho - \frac{TE_n}{\overline{TE}}\Delta\kappa\]
\[\small S_{SPC} \approx \Delta\rho - \frac{TE_n}{\overline{TE}}\Delta\kappa\]
\[\small S_{SPC} \approx\Delta\rho \]
\[\small S_{SPC} \approx \frac{TE_n}{\overline{TE}}\Delta\kappa=TE_n\Delta R_2^{\star} \]
$\Delta\kappa$ is dependent from the TE, while $\Delta\rho$ is independent (Kundu et al., 2012)
Independent Component Analysis can be use to decompose a matrix in ortogonal spatial components (with relative timecourse) FSL MELODIC webpage
A widely accepted denoising approach is ICA denoise (Griffanti et al., 2014, 2017, Pruim et al., 2015a, 2015b, Salimi-Khorshidi et al., 2014)
It is possible to use the content of $\Delta\kappa$ (and $\Delta\rho$) to estimate the number of and classify ICA components
(Kundu et al., 2012, 2013)
\[\small S_{SPC} \approx \frac{TE_n}{\overline{TE}}\Delta\kappa \qquad\qquad S_{SPC} \approx\Delta\rho \]
\[\small \alpha_v = \sum \Delta S^2_{TE_n} \]
\[\small \kappa = \frac{\sum{\alpha_v F_v,\Delta \kappa}}{\sum \alpha_v} \qquad \qquad \rho = \frac{\sum{\alpha_v F_v,\Delta \rho}}{\sum \alpha_v} \]
$F_v$ is the result of a voxelwise F-test between the residual of the fit of a model to the residual of the null model ($\alpha_v$)
- Preproc:
- (Slice timing, despiking, motion realignment)
- Brain Extraction
- Optimal Combination
- P-PCA or ted-PCA:
- PCA decomposition
- Component selection: variance or $\kappa / \rho$ threshold
- ted-ICA:
- fast-ICA (MELODIC) decomposition
- $\kappa$ and $\rho$ rank
- Decision tree
- GLM to remove bad components
The components are sorted by their $\kappa$ value (and by $\rho$), $\kappa$ and $\rho$ thresholds selected by the "elbow" method
(Kundu et al., 2012, 2013)Kundu et al., 2013: use these for ME-ICR
Current approach: use these for denoise
Take home message n. 3
We can combine ICA and ME-fMRI to automatically* denoise our data!
*more or less- The original algorithm "misfire":
- It's probably due to the original decision tree
- There's a group of volunteers openly developing and mantaining a new tedana version
- The linear model is an approximation:
- We introduced approximations twice, but most importantly, we "forgot" about the noise term, thus not separating it
- The model works only with "whole brain" analysis
- The monoexponential model is an assumption (actually two):
- The brain has different compartments
- Biexponential models reveal non-linear IV/ TE dependence(Havlicek et al., 2017)
- Data seem to be more correlated to motion after ME-ICA:
- ICA detects and removes "local" sources, thus global sources "arise"
- Solution 1: GODEC or Robust PCA(Power et al., 2018)
- Solution 2: ME-SPFM(Caballero-Gaudes et al., 2019)
- Spatial ICA might not be the best candidate to model noise:
- It assumes noise being spatially independent from BOLD
- Bayesian estimation based on "shape" of fluctuation(Cai et al., 2019)
Biexponential models reveal non-linear IV/ TE dependence(Havlicek et al., 2017)
GODEC can further clean the timeseries(Power et al., 2018)
Bayesian estimation of signals(Cai et al., 2019)
- With ME-fMRI data we can use a mathematical model of $T_2^{\star}$ decay to disentangle BOLD ($R_2^{\star}$) and non-BOLD ($S_0$) fluctuation
- We can Optimally Combine our data to get better signal
- We can use ME-ICA to further clean our data...
- ...and/or we can use other models to clean/analyse our data.
The community (at least, OHBM's) is starting to pay attention to ME-fMRI!
That's all, folks!
Thank you for your (sustained) attention!
Any question?